Installation, 4 how to install a device with a flange connection – KROHNE OPTIFLEX 2200 C_F EN User Manual

Page 39

Advertising
background image

INSTALLATION

3

39

OPTIFLEX 2200 C/F

www.krohne.com

07/2013 - 4000668002 - HB OPTIFLEX 2200 R02 en

• Use a 2.5 mm Allen wrench to attach and tighten a HC M5×5 screw (lock screw) to the bottom

segment of tube.

3.8.4 How to install a device with a flange connection

Equipment needed:

• Device
• Gasket (not supplied)
• Wrench (not supplied)

• Make sure the flange on the nozzle is level.
• Make sure that you use the applicable gasket for the flange and the process.
• Align the gasket correctly on the flange facing of the nozzle.
• Lower the probe carefully into the tank.

i

For more data on cable probes, refer to

How to install a cable probe in the tank

on page 41.

• Tighten the flange bolts.

i

Refer to local rules and regulations for the correct torque to apply to the bolts.

Figure 3-32: How to assemble the segmented coaxial probe: part 3

CAUTION!
If the lock screw is not tight, the device will not measure correctly.

Figure 3-33: Flange connection

Advertising