2 measuring pressure with external modules, 3 measuring and sourcing current (4 ma), Pressure module adapter pressure module valve – Martel Electronics DPC300A User Manual

Page 18

Advertising
background image

16

16

4.2 Measuring pressure with external modules

The calibrator has a digital interface for external pressure modules.
These modules are available for various ranges, including gauge,
vacuum, differential and absolute pressure. The modules work
seamlessly in conjunction with the calibrator. Simply plug them into
the interface and select [EXT] (external sensor). Since the interface
between the calibrator and the module is digital, the accuracy and
display resolution is dependent on the module.

REGULATOR

MAX 120 PSI

AIR SUPPLY

0-100 PSI

OUTPUT

-12-300 PSI

PUMP

-12-300 PSI

MEASURE

FINE ADJUST / VENT

PUMP

SELECTOR

MAIN

SELECTOR

WARNING:

VENT BEFORE SELECTION

HOME

PUMP

ZERO

J1

J2

J3

COM

VOLTS /mA

SWITCH TEST +24v DC

PRESSURE

MODULE

F1

F2

F3

FILTER

CHARGE

PRESSURE

VACUUM

P2

P3

P4

MARTEL ELECTRONICS

DPC-300A

P1

P2

P3

P4

Pressure Module

Adapter

Pressure Module

Valve

4.3 Measuring and sourcing current (4 ... 20 mA)

1. To measure current, use the input terminals on the front of the

calibrator. Select the mA function on the lower display. Current is
measured in mA and percentage of the measuring range. The
measuring range on the calibrator is set to 0 % at 4 mA and 100 %
at 20 mA.

Example
If the current measured is displayed as 75 %, then the value is 16
mA.
Note: The display will indicate "OL" when the measured

current exceeds the nominal range of current measurement

(24 mA).
2. The same connections are used for the current source. Select

mA-Source or mA Sim-2W from the configuration display.

Advertising