4.6.2 Calibrating a pressure-to-current transmitter
To calibrate a pressure-to-current transmitter (P/I), perform the
following:
1. Connect the calibrator and the pump to the transmitter.
2. Apply pressure with the pump.
3. Measure the current output of the transmitter.
4. Ensure that the read value is correct. If it isn't, the transmitter must

be adjusted.

REGULATOR

MAX 120 PSI

AIR SUPPLY

0-100 PSI

OUTPUT

-12-300 PSI

PUMP

-12-300 PSI

MEASURE

FINE ADJUST / VENT

PUMP

SELECTOR

MAIN

SELECTOR

WARNING:

VENT BEFORE SELECTION

HOME

PUMP

ZERO

J1

J2

J3

COM

VOLTS /mA

SWITCH TEST +24v DC

PRESSURE

MODULE

F1

F2

F3

FILTER

CHARGE

PRESSURE

VACUUM

P2

P3

P4

Use low volume

hose if possible

MARTEL ELECTRONICS

DPC-300A

P1

P2

P3

P4

Figure 5

4.6.3 Calibrating a pressure-to-voltage transmitter
To calibrate a pressure-to-voltage transmitter (P/V), perform the
following:
1. Connect the calibrator and the pump to the transmitter.
2. Apply pressure with the pump.
3. Connect the 24V to the transmitter.
3. Measure the voltage output of the transmitter.
4. Ensure that the read value is correct. If it isn't, the transmitter must

be adjusted.