000 ma, 4 measuring voltage, Test item 4 ma – Martel Electronics DPC300A User Manual

Page 19: 4 ma, Test item power supply, Dc 24 v, Test item

Advertising
background image

17

17

3. This selection can only be made in the lower display. Also, in

source mode, the calibrator will generate 0 ... 24 mA using its own
internal 24 V supply, whereas in simulation mode the calibrator
acts as a 2-wire transmitter and requires an external 24 V supply.

4. Pressing any of the arrow keys will start the output mode and

enable you to use the arrow keys to adjust the mA output. The
function keys can also be used to progress the output in either 25
% steps (4, 8, 12, 16, 20 mA) or 0 % (4 mA) and 100 % (20 mA).

5. While in the mA output mode, if the loop is broken or the

resistance burden is exceeded, the calibrator will flash "OL".

PUMP

Test Item

4...20 mA

mA MEASURE

CONFIG LOWER

MORE

16.000 mA

75.00%

J1

J2

J3

COM

VOLTS /mA

SWITCH TEST +24v DC

PUMP

Test Item

4...20 mA

mA SOURCE

MENU

LOWER

LIGHT

20.000 mA

100.00%

J1

J2

J3

COM

VOLTS /mA

SWITCH TEST +24v DC

J1

J2

J3

COM

VOLTS /mA

SWITCH TEST +24v DC

PUMP

4...20 mA

mA SIM-2W

MENU

LOWER

LIGHT

20.000 mA

100.00%

Test Item

Power Supply

+

+

J1

J2

J3

COM

VOLTS /mA

SWITCH TEST +24v DC

DC 24 V

PUMP

4...20 mA

mA MEAS/24V

MENU

LOWER

LIGHT

20.000 mA

100.00%

Test Item

4.4 Measuring voltage

For voltage measurement, use the connections on the front of the
calibrator. Select the VOLTS function on one of the displays. The
calibrator can measure up to DC 30 V.