000 v, 5 performing a pressure switch test, Test item up to dc 30 v – Martel Electronics DPC300A User Manual

Page 20: Volts config lower more, Use low volume hose if possible pressure switch

Advertising
background image

18

18

Note: The display will indicate "OL" when the measured

voltage exceeds the nominal range of voltage measurement

(30 V).

PUMP

Test Item

Up to DC 30 V

VOLTS

CONFIG LOWER

MORE

30.000 V

J1

J2

J3

COM

VOLTS /mA

SWITCH TEST +24v DC

4.5 Performing a Pressure Switch Test

REGULATOR

MAX 120 PSI

AIR SUPPLY

0-100 PSI

OUTPUT

-12-300 PSI

PUMP

-12-300 PSI

MEASURE

FINE ADJUST / VENT

PUMP

SELECTOR

MAIN

SELECTOR

WARNING:

VENT BEFORE SELECTION

HOME

PUMP

ZERO

J1

J2

J3

COM

VOLTS /mA

SWITCH TEST +24v DC

PRESSURE

MODULE

F1

F2

F3

FILTER

CHARGE

PRESSURE

VACUUM

P2

P3

P4

MARTEL ELECTRONICS

DPC-300A

P1

P2

P3

P4

Use low volume

hose if possible

Pressure

Switch

Figure 4: DPC-300A with pressure switch

To perform a switch test, the following steps must be taken:
1. The upper display is set to [P1] ST, all other displays are switched

off.

Note: The pressure switch test can be performed with the

following functions [P1] ST or EXT ST.
2. Connect the calibrator to the pressure switch using the switch

terminals. The polarity of the terminals does not matter. Then
connect the pump to the calibrator and the pressure switch.

3. Open the vent button on the pump and zero the calibrator. Then

close the vent after resetting the calibrator.

4. The top of the display will indicate "CLOSE".