9. Select the "NEW TEST" option to clear the data and perform

another test.

10. Select the "DONE" option to end the test and return to the

standard pressure setting.

Example:
[P1] ST will return to [P1].
The previous example uses a normally closed switch. The

basic procedure is effectively the same for a normally open

switch, the display will simply read "OPEN" instead of

"CLOSE".

4.6 Calibrating transmitters

4.6.1 Using the mA measurement function
The mA function enables the user to read out the 4 ... 20 mA output
from the device being calibrated. This can be done in two ways.
1. Passively – the device under test generates 4 ... 20 mA directly.

This can be read on the calibrator.

2. Actively – the calibrator supplies a loop power of DC 24 V to the

device under test to

3. Power the device while reading the resulting 4 ... 20 mA signal.

[P1]

CONFIG LOWER

MORE

psi

30.000

mA MEASURE

100.00%

20.000 mA

Passive

[P1]

CONFIG LOWER

MORE

psi

30.000

mA MEAS/24V

100.00%

20.000 mA

Active