5 remote commands and error codes – Martel Electronics DPC300A User Manual

Page 32

Advertising
background image

30

6.5 Remote Commands and Error Codes

The following tables list all commands, and their descriptions, that
are accepted by the calibrator.

Table 5: Common Commands

Command

Description

*CLS

(Clear status.) Clears the error queue.

*IDN?

Identification query. Returns the manufacturer, model num-
ber, and firmware revision level of the Calibrator.

*RST

Resets the calibrator to the power up state.

Table 6: Calibrator Commands

Command

Description

DAMP

Turns Damp on or off.

DAMP?

Returns if DAMP is on/off

DISPLAY

Turns on/off the displays specified in the command

DISPLAY?

Returns which displays are on/off

ERROR_LOOP

Turns loop power on or off in percent error mode

ERROR_LOOP?

Returns the current state of loop power in error mode

ERROR_MODE

Turns percent error mode on or off

ERROR_MODE? Returns whether percent error mode is on or off
ERROR_PORT

Set the pressure port for percent error mode

ERROR_PORT?

Returns the pressure port for percent error mode

FAULT?

Returns the most recent error code

FUNC

Sets the display mode as specified in the command

FUNC?

Returns the current mode of the upper, middle, and lower
display

HART_ON

Turns the hart resistor on.

HART_OFF

Turns the hart resistor off.

HART?

Returns the current state of the Hart resistor.

HI_ERR

Sets the 100% of span limit for percent error mode

HI_ERR?

Returns the 100% of span limit for percent error mode

IO_STATE

Set the calibrator's mA state.

IO_STATE?

Return the calibrator's mA state.

LOCAL

Returns user to manual operation of the calibrator

30

Advertising