6 entering commands – Martel Electronics DPC300A User Manual

Page 34

Advertising
background image

32

PERCENT

Percent

P1

P1 pressure measurement function

ST_P1

Switchtest mode with P1

ST_EXT

Switchtest mode with external module

SOURCE

Source state

SIM

Simulate state

UPPER

Designates Upper display

V

Voltage

Table 8: Error Codes

Error Number Error Description

100

A non-numeric entry was received where it should be a
numeric entry

101

Too many digits entered

102

Invalid units or parameter value received

103

Entry is above the upper limit of the allowable range

104

Entry is below the lower limit of the allowable range

105

A required command parameter was missing

106

An invalid command parameter was received

107

Pressure not selected

108

Invalid sensor type

109

Pressure module not connected

110

An unknown command was received

111

Bad Parameter received

112

The serial input buffer overflowed

113

Too many entries in the command line

114

The serial output buffer overflowed

6.6 Entering Commands

Commands for the calibrator may be entered in upper or lower case.
There is at least one space required between the command and
parameter, all other spaces are optional. Almost all commands for
the calibrator are sequential; any overlapped commands will be
indicated as such. This section will briefly explain each of the
commands and describe their general use, which will include any
parameters that may be entered with the command as well as what
the output of the command is.