LO_ERR?
Returns the 0% point for the percent error mode calculation.
For example:
If the 0% point is set to 20 psi, LO_ERR? would return
2.000000E+01, PSI .
MOTOR_ON
Turns the motor on.
MOTOR_OFF
Turns the motor off.
MOTOR?
Returns the state of the motor.
For example:
If the motor was on MOTOR? Would return ON.
OUT
This command also switches the calibrator into mA output
mode. A number and a unit must be entered after the command.
For example:
OUT 5 MA sets the current output at 5 mA
OUT?
Returns the output of the calibrator.
Using the above example, OUT? Would return: 5.000000E-03, A
PRES_UNIT
Used to set the pressure unit for the indicated display
For example:
To set the pressure unit to psi on the lower display send
PRES_UNIT LOWER, PSI.
PRES_UNIT?
Returns the pressure unit used when measuring pressure for
each of the 3 displays.