-18-

2)

GAIN ADJ (sensitivity adjustment)
In the following procedure, both small and large masters are used to match the
small and large master values with the bar display range.

(a) With “GAIN ADJ” displayed, press the ENT key once.

(b) The measuring unit that the item selected in 6.4.2 1) (e) uses, as well as the

current GAIN ADJ setting value, will be shown in the eight-digit display.

(c) If the GAIN ADJ setting value is not shown, it means that the item selected

in 6.4.2 1) (e) has data for two measuring units. Use the right and left arrow
keys to toggle between the data for these two measuring units.

(d) Alternately set the small master and the large master, and use the up and

down arrow keys to adjust each item's measurement unit GAIN ADJ
setting so that it approaches the small master setting and the large master
setting.

(e) After these settings are complete, press the ENT key.

(f)

“MAG ADJ” will be displayed. That concludes the GAIN ADJ setting
input procedure.

(g) Proceed to input settings for MAG ADJ.

3)

MAG ADJ (fine sensitivity adjustment)
The following procedure is used to set the coefficient for each item when one
item has two measuring units. This should be set to 1,000 for most purposes.

(a) With “MAG ADJ” displayed, press the ENT key once.

(b) The name of the measuring unit that the item selected in 6.4.2 1) (e) uses

will be shown in the eight-digit display, along with its current MAG ADJ
setting.

(c) If the MAG ADJ setting is not displayed, it is because the item selected in

6.4.2 1) (e) has data for two measuring units. Use the left and right arrow
keys toggle between the data for each one.

(d) Use the up and down arrow keys set the GAIN ADJ for each measuring

unit.

(e) After these settings are complete, press ENT.

(f)

“MIN M.” will be displayed and the unit will switch to 6.4.1 “Master
Matching.” This concludes the MAG ADJ setting procedure.