2 input from an external button, 1 overview, 2 preparation – SHIMPO CEG-2000 User Manual
Page 26
-23-
7.2
Input from an external button
7.2.1
Overview
With this unit, the following can be done from external buttons.
(1) Measuring
(2) Resetting
(3) Large master matching
(4) Small master matching
7.2.2
Preparation
SW, ETC port on the back of the unit is the port for external buttons. The D-sub 15P
(male) cable can be connected to this unit’s SW, ETC port.
7.2.3
The port’s pin layout
7.2.4
Operation with external buttons
(1) Measure button
(a) In Measurement Mode, press the Measure button.
(b) This will hold the measured value..
(2) Reset button
This releases the hold.
(3) Large Master Matching button
(a) In large master matching waiting status, set the large master to the detector.
(b) After the bar display stabilizes, press the Large Master Matching button.
(c) The unit will carry out large master matching and return to Measurement
Mode.
(d) If there is a master matching error (see 6.4.1 10)), please cope with it using
this unit’s panel switches.
(4) Small Master Matching button
(a) In Measurement Mode, set the small master to the detector.
(b) After the bar display stabilizes, press the Small Master Matching button.
(c) The unit will carry out small master matching and return to large master
matching status.
Description
Measure button
Reset button
GND
No.
1
2
3
4
5
6
7
8
Description
Large master button
Small master button
No.
9
10
11
12
13
14
15
The pin layout of the SW, ETC port on the back of this unit.
! NOTE
If master matching was unsuccessful, this data cannot be held.
! NOTE
In Set-up Mode, please choose ALL from the Selection for mask
matching (refer to 6.3.3).