7.2

Input from an external button

7.2.1

Overview

With this unit, the following can be done from external buttons.
(1) Measuring
(2) Resetting
(3) Large master matching
(4) Small master matching

7.2.2

Preparation

SW, ETC port on the back of the unit is the port for external buttons. The D-sub 15P
(male) cable can be connected to this unit’s SW, ETC port.

7.2.3

The port’s pin layout

7.2.4

Operation with external buttons

(1) Measure button

(a) In Measurement Mode, press the Measure button.
(b) This will hold the measured value..

(2) Reset button

This releases the hold.

(3) Large Master Matching button

(a) In large master matching waiting status, set the large master to the detector.
(b) After the bar display stabilizes, press the Large Master Matching button.
(c) The unit will carry out large master matching and return to Measurement

Mode.

(d) If there is a master matching error (see 6.4.1 10)), please cope with it using

this unit’s panel switches.

(4) Small Master Matching button

(a) In Measurement Mode, set the small master to the detector.
(b) After the bar display stabilizes, press the Small Master Matching button.
(c) The unit will carry out small master matching and return to large master

matching status.

Description

Measure button

Reset button

GND

No.

1
2

3

4

5

6

7

8

Description

Large master button

Small master button

No.

9

10

11

12

13

14

15

The pin layout of the SW, ETC port on the back of this unit.

! NOTE

If master matching was unsuccessful, this data cannot be held.

! NOTE

In Set-up Mode, please choose ALL from the Selection for mask
matching (refer to 6.3.3).