11 Breakdowns: Causes and Responses

12 Points of Caution

(1) Electrical wiring

The power cable included standard with this unit is 100V. To use the unit with voltage that
exceeds 125V, please use the 250V power cable.

(2) Do not use a hard tool, something sharp, to press the operation keys.
(3) About exporting these units

Export of this product may fall under the Export Control Directive, and may therefor require
the permission of the Ministry of Trade and Industry. Please discuss this with a
representative at our nearest sales office.

(4) Other

This product’s specifications may be changed without prior notice.

Symptom

Location of breakdown or

maladjustment

Response

Repeat precision is
unstable

Movement of the part that secures the
measuring unit

Tighten the loose part.

The bar graph
display does not
move

1) The proper amount of power is not

being supplied.

2) The detector is maladjusted.
3) The unit is in Hold Measurement

Mode (the bar graph is green or dark
red).

4) The unit is in Set-up Mode (orange).

1) Supply power in the 85-264V range.
2) Adjust the detector.
3) Release using the RST (reset) key.
4) Quit Set-up Mode.

The bar graph and
digital display are
blinking.

1) Power is not being supplied.
2) A fuse has blown.
3) The power supply or an internal

circuit has failed.

1) Supply power in the 85-264V range.
2) Replace the fuse (3A).
3) Request service from the

manufacturer.