11

2.4 WIRING CONNECTIONS

WARNING: Avoid electrical shock. AC power wiring must not be connected at the source distribution panel until all wiring
connections are completed.

All wiring connections are typically made to the instrument at the time of installation. All connections are made behind the
access door on the recorder right side. Loosen the two thumbscrews and open the access door to complete all wiring. Con-
nections should be made at the terminal blocks , two 14 gauge wires maximum, copper conductors only, except thermocouple
inputs. Terminal blocks are designated TB1 through TB4. See Figure 2-4 for the terminal block locations. The number of
terminal blocks present on the instrument depends on the model/hardware configuration.

FIGURE 2-4

1

1

1

1

TB1

TB2

TB3

TB4

JU1

JU2

Power Wiring

Pen 1 Input

Pen 2 Input

Communications

Recorder

Platen

Terminals located

behind access door