Badger Meter MR Transmitter User Manual

Page 16

Advertising
background image

Password

Password protection prevents unauthorized users from changing programming information . Initially the password is set to all

zeros . Its displayed name is PASSWORD and is viewed or changed using the Numeric Value Entry Procedure .

Restore Defaults

This feature allows you to restore factory calibration data . Its displayed name is RES DFLT . To restore factory calibration data,

select YES, then press ENTER .

Advanced Programming Descriptions

Advanced programming allows the user access to re-configure the analog output . Calibration of the analog output is preset

at the factory, but can be changed to customize calibration for your installation .
To access the Advanced Programming Options, press and hold MENU for approximately 3 seconds until DISPLAY is viewed on

the display panel . The programming menus will begin with display mode DISPLAY and continue as described above through

output mode OUT MODE .
After output mode has been entered, Advanced Programming starts with the following:

Calibration of Analog Output

This selection allows access to the calibration and testing of the analog output signal .
To test or change the analog output calibration, it is first necessary to change the default setting for CAL OUT? from NO to YES .

OTEE:

N

Setup prompts and descriptors for configuring and calibrating the analog output will correspond to the output

mode selected . Refer to the Flow Chart .

1 . At the CAL OUT? prompt press ENTER . NO will display .
2 . To change to YES, press either arrow key .
3 . The analog output will go to its minimum output level . A numeric value between 0-4000 will display . This is an internal

number used to drive the analog output .

4 . To increase the analog output signal level, press

. To decrease the analog output signal level, press

.

5 . Press ENTER to store the setting .
6 . The analog output will go to its maximum output level . A numeric value between 0-4000 will display . This is an internal

number used to drive the analog output .

7 . To increase the analog output signal level, press

. To decrease the analog output signal level, press

.

8 . Press ENTER to store the setting .
9 . The unit will advance to the analog output test mode . The analog output will go to its minimum output level . A numeric

value of 0 will display . For test purposes, the analog output signal can be run up or down in increments of 1 milliamp or 1

volt, depending on the OUT MODE selected .

10 . To increase the analog output signal level, press

. To decrease the analog output signal level, press

.

11 . Press ENTER to exit the analog calibration mode .
12 . The unit automatically advances to the PASSWORD feature .

Password

Password protection prevents unauthorized users from changing programming information . Initially the password is set to all

zeros . Its displayed name is PASSWORD and is viewed or changed using the Numeric Value Entry Procedure f .

Restore Defaults

This feature allows you to restore factory calibration data . Its displayed name is RES DFLT . To restore factory calibration data,

select YES, then press ENTER .