Maintenance, Cartridge cleaning, Inspection – Badger Meter MR Transmitter User Manual

Page 17: Cartridge cleaning inspection

Advertising
background image

MAINTENANCE

BEFORE ATTEMPTING TO REMOVE THE TRANSMITTER FROM THE LINE, CHECK THE SYSTEM TO CONFIRM THAT LINE

PRESSURE HAS BEEN REDUCED TO ZERO PSI. FAILURE TO FOLLOW THESE INSTRUCTIONS COULD RESULT IN SERIOUS

PERSONAL INJURY OR DEATH AND/OR DAMAGE TO THE EQUIPMENT.

DISCONNECT ELECTRICAL POWER BEFORE REMOVING METER COVER. FAILURE TO FOLLOW THESE INSTRUCTIONS

COULD RESULT IN SERIOUS PERSONAL INJURY OR DEATH AND/OR DAMAGE TO THE EQUIPMENT.

Cartridge Cleaning

1 . Disconnect the transmitter cable .
2 . Remove the meter from the line . Remove excess piping from the transmitter . See

Figure 4

and

Figure 5

.

OTEE:

N

It is not necessary to remove the aluminum housing from the transmitter to remove it from the line .

3 . Thoroughly wipe off the entire transmitter surface using mild detergent or isopropyl alcohol .

DO NOT USE AROMATIC HYDROCARBONS, HALOGENATED HYDROCARBONS, KETONES OR ESTER BASED FLUIDS ON

POLYCARBONATE LENS. FAILURE TO FOLLOW THESE INSTRUCTIONS COULD RESULT IN DAMAGE TO THE TRANSMITTER.
4 . Remove the inlet port cap, wave spring, retaining ring, and cone assembly from the transmitter body (

Figure 9

) .

5 . Gently push the body towards the outlet port .
6 . The piston, inner magnet and transmitter spring are secured within the transmitter body with a retaining ring . Remove the

retaining ring with a small screwdriver, then the internal components can be removed from the body (

Figure 9

) .

OTEE:

N

If internal parts do not slide freely from cartridge, use a wooden dowel inserted into the outlet port of the meter to

push parts out .

7 . Place all parts on a clean work surface . Clean and inspect all parts . Replace any that appear worn or damaged . Check inlet

port O-ring for damage and replace if required .

FIELD REPLACEMENT OF THE SPRING, METERING CONE AND/OR PISTON/MAGNET ASSEMBLY MAY RESULT IN CHANGES

TO THE CALIBRATION OF THE FLOW METER.
8 . Reassemble the transmitter by inserting the transmitter spring into the body, followed by the piston/inner magnet

assembly . A slight compression of the piston against the spring is required during installation of the retaining ring .

9 . Gently push body assembly into the outlet end of the transmitter enclosure . The flat surface of the body output port

should be flush with the transmitter enclosure opening .

10 . With the transmitter positioned vertically on a flat surface, inlet port facing up, install the transmitter cone assembly and

wave spring into the body and secure with the inlet port end cap .

11 . Reinstall transmitter to the line . Reconnect electrical power .

Inspection

1 . Frequent inspection should be made . The environment and frequency of use should determine a schedule for

maintenance checks . It is recommended that it should be at least once a year .

2 . Perform visual, electrical, and mechanical checks on all components on a regular basis .
3 . Visually check for undue heating evidence such as discoloration of wires or other components, damaged or worn parts, or

leakage evidence such as water or corrosion in the interior .

4 . Make sure all electrical connections are clean and tight and that the device is wired properly .

User Manual

Page 17

November 2013

Advertising
This manual is related to the following products: