Manual Speed Adjust

OFF

Manual Speed Adjust

Press START to run

4.3.2

4 - 20 mA Input

Used to remotely control pump with an incoming 4-20 mA signal.

Default settings:

4 mA = 0% motor speed
20 mA = 100% motor speed

Press SELECT RUN MODE button until 4 - 20 mA Input is displayed in the top
line of the display.

To configure the pump, navigate to 4 - 20 mA Input menu by using short-cut
method described at beginning of section, or by pressing MENU button, then
selecting Input Setup, Input Modes, and then 2 - 20 mA Input.

Press UP or DOWN arrow to scroll through 0 - 9 on selected digit.

Press RIGHT arrow to scroll over to next digit to right. If you pass your desired
digit, you can easily scroll back by continuously pressing RIGHT button.

Press ENTER to save changes.

Continue this process until all four screens have been configured.

If you used short-cut to enter 4-20 mA input setup, then just press and hold
Select Run Mode button until Run Mode screen is displayed.

If you used Menu button to navigate to 4-20 mA input setup, you must navigate
back out of menu structure. To do this you must select <-Esc at bottom of every
screen menu until you see Run Mode screen displayed.

- 4-20 mA Input

% Motor @ Min Signal:

Input % Motor Speed at
Minimum mA Signal

00.0%

0

- 4-20 mA Input

Min mA Signal:

Input Min. mA signal Input.

4.00

0

- 4-20 mA Input

% Motor @ Max Signal:

Input % Motor Speed at
Maximum mA Signal

00.0%

1

- 4-20 mA Input

Max mA Signal:

Input Max. mA Signal Input

0.00

2

Press ENTER to set.

Press ENTER to set.

In 4 - 20 mA mode, you can view current
incoming signal by pressing RIGHT arrow (see
screen shot). RIGHT arrow is a convenient
way to scroll through multiple read-only
screens during normal operating mode of
pump.

Tip!

4-20 mA Input

mA

Displays current incoming signal

12.5

- Screen Shot -

Example 1

6

4

8 10 12 14 16 18 20

0

25

50

75

100

Pump

Motor Speed (%)

Milliamp input (mA)

from external source

4 mA = 0% Pump Output

20 mA = 100% Pump Output

Example 2

6

4

8 10 12 14 16 18 20

0

25

50

75

100

4 mA = 100% Pump Output

20 mA = 0% Pump Output

Example 3

6

4

8 10 12 14 16 18 20

0

25

50

75

100

4 mA = 20% Pump Output

16 mA = 43.8% Pump Output

Е

Й

ДИ

Displays current rotor direction

0%

Output

Pump

Motor Speed (%)

Pump

Motor Speed (%)

100%

Output

100%

Output

0%

Output

43.8%

Output

20%

Output

Milliamp input (mA)

from external source

Milliamp input (mA)

from external source