Symptom Cause
Remedy
Measurement value
contains an error.
z
Inputted pipe specifications are
different from actual values.
z
Scales exist on wall of old pipe.
Difference in inner diameter of 1%
produces about 3% of error.
z
Input correctly.
z
When inputting, regard the scales
as lining.
z
Straight pipe is inadequate
(10D upstream and 5D
downstream cannot be secured).
Find a better place for mounting the
sensor (upstream the disturbance).
There must be no disturbance within
30D upstream.
Pump, valve, joint pipe, etc. are not
allowed.
z
Mount the sensor at different
angles with respect to pipe cross-
section until an average is
obtained.
z
Pipe is not filled with water or
contains mud or sand.
Reading rises as cross-sectional area
reduces.
z
Move to vertical pipe.