How to set the flow rate unit, Bulletin f-70 – Dwyer UXF2 User Manual
Page 85

- 75 -
4.6.2. How to set the flow rate unit
Description
z
Select the unit of flow rate.
z
Metric system
Flow rate···················· L/s (factory set), L/min, L/h, ML/d, m
3
/s, m
3
/min, m
3
/h, Mm
3
/d, BBL/s, BBL/min,
BBL/h, MBBL/d
z
Inch system
Flow rate···················· gal/s, gal/min, gal/h, Mgal/d, ft
3
/s, ft
3
/min, ft
3
/h, Mft
3
/d, BBL/s, BBL/min, BBL/h,
MBBL/d
<Caution> First, set the unit system (metric or inch) according to Section 4.6.1.
For actual keying, refer to the typical operation indicated below. Set the parameter protection to OFF beforehand
(Section 4.4.1).
Typical operation Set the flow rate unit to [m
3
/h].
Keying order
Description
Indication
ź
Pressed 3 times to indicate "MEASURE SETUP".
MEASURE SETUP
ENT
ź
Pressed to indicate "SYSTEM UNIT".
SYSTEM UNIT
METRIC
ź
Pressed to indicate "FLOW UNIT".
FLOW UNIT
L/S
ENT
ź
Pressed to blink.
FLOW UNIT
L/S
ź
Pressed 6 times to select "m
3
/h".
FLOW UNIT
m3/h
Pressed to register it.
FLOW UNIT
½½ COMPLETE ½½
Ļ
ENT
ź
ź
ź
ź
--- "m
3
/h" registered.---
FLOW UNIT
m3/h
ESC
Press
ESC
key once, and
key twice to resume
the measurement mode.
0.000 m/s
0.000 m3/h
BULLETIN F-70