5.7.4. If analog output is abnormal
Symptom Cause
Remedy
Current output is
wrong.
Range setting is wrong.
z
Set the range correctly.
Not 4 mA when
measurement value is 0.
Analog output is misadjusted.
z
Calibrate the analog output.
Output is 0 mA.
Cable is open-circuited.
Output is over 20 mA.
"OVERFLOW" is indicated on LCD.
Range over.
z
Recheck and set range data of
analog output.
Output is below 4 mA.
"UNDERFLOW" is indicated on
LCD.
Back flow
z
Set the upstream and downstream
properly.
Measurement value
varies but analog output
is constant.
Output load is above 600
:.
z
Lower the load below 600
:.
Analog output does not
match the measurement
value.
Analog output is misadjusted.
z
Calibrate the analog output.
Output remains
unchanged even after
analog output calibration.
Hardware error.
z
Contact manufacturer.