Calibration theory, Hard and soft iron effects, 1 calibration theory – PNI TCM 5 User Manual

Page 23: 2 hard and soft iron effects

Advertising
background image

PNI Sensor Corporation

Doc #1007537 r12

TCM3 & TCM5 User Manual

– Sept 2012

Page 23

4.2.1 Calibration Theory

The exact calibration method will depend on the actual settings of the calibration parameters. An

example of the various settings and their effect can be seen in the TCM Studio

– Evaluation

Software section.

The main object of the calibration is to allow the TCM to calibrate out any distortions to the magnetic

field caused by the host system. To that end the TCM needs to be mounted within the host system

and the entire application needs to be moved as a single unit during the calibration. Movement

should include at least 180

˚ of horizontal rotation, but to achieve the highest accuracy a full 360˚ of

horizontal rotation with as many different tilt angles as possible during the rotation is required.

To achieve the highest accuracy throughout the TCM

’s entire tilt range, the unit will need to be tilted

through the entire range. For example, if the unit is only tilted through 40

˚ of pitch and roll, then the

heading information from the TCM will only be accurate through 40

˚ of pitch and roll. For maximum

performance the TCM should be exposed to tilt angles covering a full 360°, meaning upside down.

Recommended calibration procedure for taking the minimum number of sample points

follows.

4.2.2 Hard and Soft Iron Effects

Hard iron distortions are caused by permanent magnets and magnetized steel or iron object within close

proximity to the sensors. This type of distortion will remain constant and in a fixed location relative to the

sensors for all heading orientations. Hard-iron distortions will add a constant magnitude field component

along each axis of sensor output and can be easily compensated for using a simple saturation method.

Soft-iron

distortions are the result of interactions between the Earth’s magnetic field and any

magnetically “soft” material within close proximity to the sensors. In technical terms, soft materials have a

high permeability. The permeability of a given material is a measure of how well it serves as a path for

magnetic lines of force, relative to air, which has an assigned permeability of one.

The TCM 3-axis digital compass features soft-iron and hard-iron correction.



Advertising
This manual is related to the following products: