Sensor Heating (RoHumiHeat), Time-Dependent

The sensor heating can be activated cyclically. The following time intervals are available:

Hourly

Daily

Weekly

Monthly

The duration of the heating cycle is also freely selectable in minutes.

The temperature of the heated sensor can be selected freely between 0.1 °C and 10 °C above ambient

temperature.

ROTRONIC offers the time-dependent heating function as a feature for applications in which cyclical heating

suffices. Please note, however, that condensation could form on the sensor in an unheated state. If

condensation does form on the sensor when it is in operation, it can only be removed reliably with the

RoHumiClean function (five minutes cleaning at 150 °C).

Temperature differences of more than 3 °C between the sensor and environs caused by heating can have

effects on the external temperature measurement. This is particularly possible in stagnant air because the

heat energy is then not dissipated efficiently.

The start and stop responses follow the general function description above. The function "time-dependent

sensor heating" can be enabled/disabled with the HW4 software.