Draft, E-m-hm4-v0_08 – ROTRONIC HM4 User Manual

Page 36

Advertising
background image

E-M-HM4-V0_08

ROTRONIC AG

Bassersdorf, Switzerland

Document Code

Unit

HM4 Humidity and Temperature Probe with

Heated Sensor

Instruction Manual

Document Type

Page 36 of 41

Document Title

© 2014; ROTRONIC AG

E-M-HM4-V0_08

Digital Interface

Interface type

RS-485

Baud rate

: 19,200

Parity

: none

Data bits

: 8

Stop bits

: 1

Service Connector

Interface type

UART (TTL Level), inside housing

Max. length of service cable

5 m (16.4 ft)

Sensor Heating (RoHumiHeat)

Heating temperature

0.1…10 °C above ambient temperature

Operating modes

Permanent, time interval or value-dependent

Sensor Cleaning (RoHumiClean)

Heating temperature

To 150 °C

Operating modes

Programmable time interval

General Specifications

Probe material

POM (

polyoxymethylene)

Filter material

Polyethylene, 20 µm pore size

Dimensions

238 x 29 mm (model with connector at housing)

281 x 25 mm, plus cable depending on order code (model with cable)

Weight

150 g (model with connector at housing)

Conformity with Standards

CE / EMC immunity

EMC Directive 2004/108/EC: EN 61000-6-1: 2007, EN 61000-6-2: 2005

EN 61000-6-3: 2007, EN 61000-6-4: 2007

Solder type

Lead free (RoHS directive)

Fire protection class

Corresponds to UL94-HB

FDA / GAMP directives

CFR21 Part 11 and GAMP5

DRAFT