HM4 Humidity and Temperature Probe with

Heated Sensor

Instruction Manual

Document Type

Page 18 of 41

Document Title

© 2014; ROTRONIC AG

E-M-HM4-V0_08

4.1.6

Sensor Cleaning (RoHumiClean)

Sensor cleaning can take place in three time intervals.

Daily

Weekly

Monthly

The duration of the cleaning cycle is freely selectable in minutes. The cleaning temperature can be set up to

150 °C. No measurements are carried out during the cleaning phase. The last measured value is frozen until

the cleaning cycle is finished.

Stop Response

After cleaning, the sensor needs to cool down. No values are measured for a time of five minutes after

cleaning until this state has been reached.

Working Principle

Heating of the sensor up to 150 °C removes volatile and non-volatile contaminants. The sensor is thus cleaned

and, if water has entered the housing, dried completely. This prevents electrochemical corrosion.

Note:

When the sensor is heated to 150 °C, a drop of water on the surface of the sensor evaporates in about three

minutes.

A minimum power supply of 15 VDC is needed to initiate sensor cleaning manually with the HW4 software. If

this is not available in the field application, the function is still activated, but might not reach the required

temperature. The limit values are as follows:

Power supply: 15 VDC attainable heating temperature: 150 °C

Power supply: 10 VDC attainable heating temperature: 100 °C

Power supply: 5 VDC attainable heating temperature: 35 °C

The function can be enabled/disabled with the HW4 software.