6.2 Wiring

This section describes the wiring of the device with connection possibilities.

6.2.1

Electrical Diagrams

The HM4 is connected with the 7-pin Tuchel

connector or the connection cable with open ends.

The maximum permissible cable length depends

on the voltage drop caused by the flow of current

to the devices connected to the HM4 outputs. The

effective load of a HM4 output comprises the cable

and load resistance and should amount to at least

1000

Ω/V (for voltage outputs according to Figure

3). The cable resistance should not be more than

1/1000 of the load resistance.

Figure 3: Voltage outputs

The maximum permissible length of the cable

connecting the unit to other devices is determined

by the total resistance resulting from addition of

the cable resistance and resistance of the devices

connected to the unit in series. This resistance

should not exceed 500 ohm (load) (for current

outputs according to Figure 4).

Figure 4: Current outputs