1.1 Scope

This manual is valid for the HM4 probe series with firmware version V1.1. The low-order digit of the firmware

version stands for minor changes, e.g. correction of errors, that do not influence the main functionality of the

device.

1.2 Background Information on the Principle of Measurement

In environments with high humidity (>90 %RH) near saturation point there is a risk of condensation forming on

the humidity sensor. To avoid resultant incorrect measurement, the sensor can be heated above ambient

temperature to avoid condensation. The same applies for outdoor measurements in mist or high levels of dew.

The probes in the HygroMet4 series from ROTRONIC (HM4) measure temperature and humidity

simultaneously. Formation of condensation on the capacitive SMD humidity sensor can be prevented by raising

the temperature above ambient temperature, thereby enabling accurate humidity measurements even when the

measurement environment has already reached dew point (100 %RH).

Pt100 temperature sensor

Inside: Evaluation electronics with AirChip3000 and real-time clock

Heated SMD Thermo humidity sensor

DRAFT