Dew point accuracy, Draft – ROTRONIC HM4 User Manual

Page 38

Advertising
background image

E-M-HM4-V0_08

ROTRONIC AG

Bassersdorf, Switzerland

Document Code

Unit

HM4 Humidity and Temperature Probe with

Heated Sensor

Instruction Manual

Document Type

Page 38 of 41

Document Title

© 2014; ROTRONIC AG

E-M-HM4-V0_08

10.1 Dew Point Accuracy

The HM4 can be configured to calculate the dew point on the basis of the measurement of the relative humidity

and temperature. The accuracy of this conversion varies depending on the humidity and temperature conditions

as shown in Figure 8:

Figure 8: Dependency of the accuracy of dew point calculation from the measured humidity and temperature
values

Example:

At a temperature of 20 °C, a dew point value between -27 °C and 20 °C is calculated with an accuracy of

±0.5 °C or better.

DRAFT

Advertising