Firmware Update

The HM4 firmware can be updated with the HW4 software. Firmware updates are available for downloading on

the ROTRONIC website. They are written to the HM4 probe with an AC3010, AC3010-T or AC3011 service

cable.

IMPORTANT:

There must be a permanent connection to the computer and a stable power supply during the update process.

Procedure:

Connect the HM4 with one of the above-mentioned connection cables or devices.

Note:

As from 2014 the service cables have a standard USB driver. Older cables need a ROTRONIC

USB driver, which is installed on the PC together with the HW4 software. See the manual E-M-

HW4v3-Main for detailed information on the ROTRONIC USB driver.

Download the latest firmware to your PC from the ROTRONIC website.

Launch the HW4 software on the PC and look for the HW4 probe.

HW4 Main Menu > Devices and Groups > Search for Devices > Search for USB Master / Search

for Ethernet Master

Under Device Manager in the device tree, select Extras > Firmware Update

For further information, see the manual E-M-HW4v3-F2-023