Sper Scientific 850046K AquaShock DO Kit User Manual
Page 12
13
Set MTC (Manual Temperature Compensation)
Temperature Coefficient
Note...
Only perform this setting when ATC Temperature Probe
is not available.
1. Press
RANGE for 2 seconds to enter the Set
Temperature Mode (see Fig. A).
The temperature value will flash (see Fig. B).
2. Press ↑ to increase the Temperature Value.
Hold down ↑ to continually increase the value.
Press ↓ to reduce the Temperature Value.
Hold down ↓ to continually reduce the value.
If the temperature unit is °C, the range for adjustment
is 0.0~50.0°C.
If the temperature unit is °F, the range for adjustment
is 32.0~122.0°F.
3. Press
SETUP to confirm change and return to Normal Mode
(see Fig. D).
Press
EXIT to return to Normal Mode without saving change
(see Fig. C).