Sper Scientific 850046K AquaShock DO Kit User Manual
Page 14

14
15
Set Barometric Pressure Parameters
1. Press
SETUP to enter the Setup Mode.
2. Press ↑ or ↓ until the Bar. Pressure screen is seen (see Fig. A).
3.
SETUP to enter the Select Bar. Unit screen (see Fig. B or Fig. C).
4. Press ↑ or ↓ to scroll through the available barometric
measurement scales.
5. Press
SETUP to save the displayed measurement scale
and enter the Set Bar. Value screen (see Fig. D or Fig. E).
The current Bar. Value number will flash.
6. Press ↑ to increase the Bar. Value.
Hold down ↑ to continually increase the value.
7. Press ↓ to reduce the Bar. Value.
Hold down ↓ to continually reduce the value.
If the Bar. Value Measurement scale is mmHg the
available range is 500~1499.
If the Bar. Value measurement scale is kPa the
available range is 66.6~199.9
8. Press
SETUP to save the changes and return to Normal Mode.
9. Press
EXIT to return to Normal Mode without saving changes.