Sper Scientific 850046K AquaShock DO Kit User Manual
Page 19
Advertising
19
100% Calibration
Note...
Only when the measurement value is above 50% can the meter be
calibrated at 100%. All other keys except
POWER ON/OFF and
HOLD are inoperable while performing calibration.
1. Clean the probe thoroughly with deionized water.
2. Press
RANGE to select the % saturation scale (see Fig. A).
3. Wrap the probe with a wet cloth being sure not to touch
the probe’s oxygen permeation membrane.
4. Press
CAL for 2 seconds to enter the calibration mode
(see Fig. B).
5. When calibration is complete the display will read
“Finished” (see Fig. C).
6. Press
EXIT to return to Normal Mode (see Fig. D).
Fig. A
Fig. D
Fig. B
Fig. C