Sper Scientific 850046K AquaShock DO Kit User Manual

Page 19

Advertising
background image

19

100% Calibration
Note...

Only when the measurement value is above 50% can the meter be

calibrated at 100%. All other keys except

POWER ON/OFF and

HOLD are inoperable while performing calibration.

1. Clean the probe thoroughly with deionized water.

2. Press

RANGE to select the % saturation scale (see Fig. A).

3. Wrap the probe with a wet cloth being sure not to touch

the probe’s oxygen permeation membrane.

4. Press

CAL for 2 seconds to enter the calibration mode

(see Fig. B).

5. When calibration is complete the display will read

“Finished” (see Fig. C).

6. Press

EXIT to return to Normal Mode (see Fig. D).