Sper Scientific 850046K AquaShock DO Kit User Manual

Page 22

Advertising
background image

22

23

View Calibration Points

1. Press

RANGE chose the scale you want view.

2. Press

SETUP to enter the Setup mode.

3. Press ↑ or ↓ to select the View Cal Data screen (see Fig. A).

4. Press

SETUP to enter the mode. (The 100% scale will display

as in Fig. B and Fig. C. The 0% scale will display as in

Fig. E and Fig. F). The date and time is a cyclic display.

If there are no calibration points set the display will

show “- - - -” (see Fig. D).

5. Press ↑ once to view the previous Cal point.

Repeat to view all previous Cal points.

Press ↓ once to view the next Cal point.

Repeat to view all additional Cal points.

6. Press

SETUP or EXIT to return to Normal Mode.

Fig. A

Fig. B

Fig. C

Fig. D