Chapter 4 ________________________________________________________________ Operation

VAISALA_______________________________________________________________________ 53

Byte 5-7:

Average wind direction (degrees) CR
LF

Bytes 8-10:

Wind direction at peak wind speed
(degrees) CR LF

Bytes 11-12:

Averaging time for wind speed and
direction CR LF

Bytes 13-14:

Averaging time for peak wind speed
and direction CR LF

Bytes 15-19:

Average wind speed CR LF

Bytes 20-24:

Peak wind speed CR LF

Byte 25:

Wind speed units (K = knots) CR LF

Byte 26-27:

Signal quality CR LF

Bytes 28-29:

Hex checksum CR LF

Byte 30:

ETX CR LF

Byte 31:

Carriage return CR LF

Byte 32

Line feed CR LF

WL

Enable automatic heater control CR LF

WM

Lock heaters on CR LF

WN

Lock heaters off CR LF

WS

Description of BIT status CR LF

WJ

Set averaging times for wind speed/direction and peak
wind speed/direction CR LF
Enter password when prompted, then selections to the
following prompts CR LF
Wind speed/direction averaging time = ? CR LF
Peak wind speed and direction averaging time = ? CR LF

WD

Perform and report extended diagnostic tests CR LF

WF

Return fixed data in WA command format CR LF

WR

Reset sensor CR LF

WCAL

Perform Sensor Self Calibration CR LF

WH

Display this help menu CR LF