SPECIFICATIONS

Power:

100 to 240 Vac ± 10 %, 50/60 Hz. Maximum consumption: 20 VA.

24 V Model: 24 Vac/dc ±10%, 50/60 Hz. Maximum consumption: 10 VA.

Environmental Conditions: Operating temperature: 0 to 50 °C. Relative Humidity: 80 % to 30 °C. For temperatures
above 30 °C, decrease 3 % per °C.

For internal use.

Installation category II.

Pollution degree II.

Altitude < 2000 m.

Dimensions: 164 x 117 x 70 mm

Weight: 400 g

Housing: ABS+PC

Protection: IP20

Analog Inputs:

The types of input signals accepted by FieldLogger and their maximum ranges of measurement are selected in the
Configuration software and are listed in the following table.

INPUT TYPE

MEASURING RANGE

ACCURACY

Thermocouple J

-120 to 1000 ºC (-184 ºF a 1832 ºF)

± 0.2 % (F.R.) ± 1 ºC

Thermocouple K

-130 to 1372 ºC (-202 ºF to 2501,6 ºF)

± 0.2 % (F.R.) ± 1 ºC

Thermocouple T

-130 to 400 ºC (-202 ºF to 752 ºF)

± 0.2 % (F.R.) ± 1 ºC

Thermocouple E

-130 a 780 ºC (-202 ºF to 1436 ºF)

± 0.2 % (F.R.) ± 1 ºC

Thermocouple N

-130 to 1300 ºC (-202 ºF to 2372 ºF)

± 0.2 % (F.R.) ± 1 ºC

Thermocouple R

20 to 1768 ºC (68 ºF to 3214,4 ºF)

± 0.2 % (F.R.) ± 3 ºC

Thermocouple S

20 to 1768 ºC (68 ºF to 3214,4 ºF)

± 0.2 % (F.R.) ± 3 ºC

Thermocouple B

100 to 1820 ºC (212 ºF to 3308 ºF)

± 0.2 % (F.R.) ± 3 ºC

Pt100

-200 to 850 ºC (-328 ºF to 1562 ºF)

± 0.15 % (F.R.)

Pt1000

-200 to 850 ºC (-328 ºF to 1562 ºF)

± 0.15 % (F.R.)

Linear 0 to 20 mA

Configurable

± 0.15 % (F.R.) *

Linear 4 to 20 mA

Configurable

± 0.15 % (F.R.) *

Linear 0 to 20 mV

Configurable

± 0.15 % (F.R.) *

Linear 0 to 50 mV

Configurable

± 0.15 % (F.R.) *

Linear 0 to 60 mV

Configurable

± 0.15 % (F.R.) *

Linear -20 to 20 mV

Configurable

± 0.15 % (F.R.) *

Linear 0 to 5 V

Configurable

± 0.15 % (F.R.) *

Linear 0 to 10 V

Configurable

± 0.15 % (F.R.) *

F.R. = Full Range = Span

(*) Note: The full scale refers to the input of the sensor signal and not the range of configured indication.

Table 04 - List of signals accepted by FieldLogger

Accuracy is guaranteed at intervals of reading greater than 0.2 seconds per channel. For faster read rates (intervals of
less than 0.2 seconds), there will be a loss of accuracy that will be as increased as the rate of reading. Likewise,
although most of the channels can indicate a little beyond the range limits set, the specification is not guaranteed outside
the range.

Input impedance of the analog channels:

Thermocouples / Pt100 / Pt1000 / mV: > 2 MΩ

mA: 15 Ω + 1.5 V

V: 1.1 MΩ

Maximum compensated Pt100/Pt1000 cable resistance: 40 ohms

Excitation current:
• Pt100s: 360 µA