Please read the following carefully!

This instrument may sometimes be used with hazardous materials, operations, and equipment.
It is beyond the scope of this manual to address all of the potential safety problems associated with its
use in such applications. It is the responsibility of the user of this pipette to consult and establish appro-
priate safety and health practices and determine the applicability of regulatory limitations prior to use.

Safety Instructions

1. Every user must read and understand this

operating manual prior to using the instrument
and observe these instructions during use.

2. Follow general instructions for hazard preven-

tion and safety instructions; e.g., wear protec-
tive clothing, eye protection and gloves.

When working with infectious or other hazard-
ous samples, all appropriate regulations and
precautions must be followed.

3. Observe all specifications provided by reagent

manufacturers.

4. Never use the instrument in an atmosphere

with a danger of explosion. Highly flammable
liquids must not be pipetted.

5. Only use the instrument for pipetting liquids

that conform to the specifications defined in
the operating exclusions and limitations
(see page 41). If in doubt, contact the manu-
facturer or supplier.

6. Always use the instrument in such a way that

neither the user nor any other person is
endangered. Avoid splashes and only use suit-
able vessel.

7. Avoid touching the tip orifices when working

with hazardous samples.

8. Never use force on the instrument!
9. Only use original spare parts. Do not attempt

to make any technical alterations. Do not
dismantle the instrument any further than is
described in the operating manual!

10. Before use check the instrument for visible

damages. If there is a sign of a potential
malfunction (e.g., piston difficult to move,
leakage), immediately stop pipetting. Consult
the 'Troubleshooting' section of this manual
(see page 70), and contact the manufacturer
if needed.

11. The original battery must not be replaced with

non rechargeable batteries or rechargeable
batteries of other manufacturers.

12. To charge the NiMH battery pack, use only the

original AC adapter.

13. The AC adapter has to be protected against

moisture and must be used only for this instru-
ment.

14. Dispose of batteries only when discharged and

according to applicable regulations.

Improper use of the instrument or the batteries (short circuit, mechanical damage, overheating, incorrect
AC adapter, etc.) can lead to battery explosion.

!

Warning!