Servicing and cleaning – BrandTech Scientific Transferpette-8/-12 electronic User Manual

Page 66

Advertising
background image

Z

Do not loosen the central guide rod (Z)!

Remove both outer Phillips screws on the nose
cone assembly.

3.

Remove screws

Pull the piston and nose cone assembly apart and
separate.

4.

Separate piston

and nose cone

assembly

Place the mounting tool on the piston nut and un-
screw the piston nut. Remove the piston nut and
pull out the piston.

5.

Remove piston

Removing of pistons for cleaning or replacing

Unscrew the manifold from the handle
(See p. 64).

1.

Unscrew manifold

Turn both closures of the manifold housing cover
90° and slide off the housing.

2.

Slide housing

Attention:

Servicing and Cleaning