Servicing and cleaning – BrandTech Scientific Transferpette-8/-12 electronic User Manual
Page 65
![background image](/manuals/464065/65/background.png)
➞
➞
➞
English
65
The seal will either remain inside the nose cone
or will stay on the piston after the nose cone is
removed.
Push the piston unit down to the bottom. Remove
the seal, inspect it and clean or replace if neces-
sary
4.
Remove seal
If required, lubricate the piston lightly with the
supplied silicone oil. Push the seal on the piston
with its flat side facing up-ward. Use the mount-
ing tool to mount and tighten the cleaned or new
nose cone.
5.
Mount seal
Reassemble the manifold and mount it to the
handle.
6.
Reassemble
manifold
Performe the reference run (rEF).
7.
Performe
reference run
The manifold must be screwed into the handle
counter-clockwise, until it audibly snaps into
place.
You must not pull the manifold downward.
Check the instrument for tightness, mobility and
accuracy (see page 63 "Servicing").
Note:
Note:
Servicing and Cleaning