Accuracy table, English, 20 °c ex – BrandTech Scientific Transferpette-8/-12 electronic User Manual
Page 59
English
Final test values related to the nominal capacity (maximum volume)
and the indicated volume steps indicated on the instrument, obtained
when instrument and distilled water are equilibrated at ambient tem-
perature (20 °C/68 °F) and with smooth operation. According to
DIN EN ISO 8655.
Volume range
Volume step
A* CV*
Increment
Recommended
µl
µl
≤ ± %
≤ %
µl
type of tip, µl
0.5 - 10
10
1.2
0.8
0.01
0.5 - 20
5
2.0
1.5
1
8.0
4.0
1 - 20
20
1.0
0.5
0.02
0.5 - 20
10
2.0
1.0
2
8.0
3.0
5 - 100
100
0.8
0.25
0.1
2 - 200
50
1.6
0.4
10
4.0
1.5
10 - 200
200
0.8
0.25
0.2
2 - 200
100
1.4
0.4
20
4.0
1.3
15 - 300
300
0.6
0.25
0.5
5 - 300
150
1.2
0.4
30
3.0
1.2
Accuracy tolerances for the Transferpette
®
-8/-12 electronic
* A
= Accuracy, CV
= Coefficient of Variation
Accuracy Table
!
20 °C
Ex
59
Testing instructions (SOPs) and a demo version of the EASYCAL
™
4.0 calibration software are available
for download at www.brand.de.