Accuracy table, English, 20 °c ex – BrandTech Scientific Transferpette-8/-12 electronic User Manual

Page 59

Advertising
background image

English

Final test values related to the nominal capacity (maximum volume)
and the indicated volume steps indicated on the instrument, obtained
when instrument and distilled water are equilibrated at ambient tem-
perature (20 °C/68 °F) and with smooth operation. According to
DIN EN ISO 8655.

Volume range

Volume step

A* CV*

Increment

Recommended

µl

µl

± %

%

µl

type of tip, µl


0.5 - 10

10

1.2

0.8

0.01

0.5 - 20

5

2.0

1.5

1

8.0

4.0

1 - 20

20

1.0

0.5

0.02

0.5 - 20

10

2.0

1.0

2

8.0

3.0


5 - 100

100

0.8

0.25

0.1

2 - 200

50

1.6

0.4

10

4.0

1.5

10 - 200

200

0.8

0.25

0.2

2 - 200

100

1.4

0.4

20

4.0

1.3

15 - 300

300

0.6

0.25

0.5

5 - 300

150

1.2

0.4

30

3.0

1.2

Accuracy tolerances for the Transferpette

®

-8/-12 electronic

* A

= Accuracy, CV

= Coefficient of Variation

Accuracy Table

!

20 °C

Ex

59

Testing instructions (SOPs) and a demo version of the EASYCAL

4.0 calibration software are available

for download at www.brand.de.