Servicing and cleaning – BrandTech Scientific Transferpette-8/-12 electronic User Manual

Page 67

Advertising
background image

English

Insert the cleaned or new piston. Screw on the
piston nut and tighten with the mounting tool

6.

Mount piston

Loosen the nose cones by a half rotation. After-
ward push the nose cone assembly on the piston
unit and fasten it. Then tighten nose cones.

7.

Reassemble piston

and nose cone

assembly

Reassemble the manifold and mount it to the
handle.

8.

Mount maifold

Performe the reference run (rEF).

9.

Performe

reference run

The manifold must be screwed counter-clockwise
into the handle so that it audibly snaps into place.
You must not pull the manifold downward.

Check the instrument for tightness, mobility and
accuracy (see page 63 "Servicing").

Note:

Servicing and Cleaning

67