Servicing and cleaning – BrandTech Scientific Transferpette-8/-12 electronic User Manual
Page 67
Advertising
![background image](/manuals/464065/67/background.png)
➞
➞
➞
English
Insert the cleaned or new piston. Screw on the
piston nut and tighten with the mounting tool
6.
Mount piston
Loosen the nose cones by a half rotation. After-
ward push the nose cone assembly on the piston
unit and fasten it. Then tighten nose cones.
7.
Reassemble piston
and nose cone
assembly
Reassemble the manifold and mount it to the
handle.
8.
Mount maifold
Performe the reference run (rEF).
9.
Performe
reference run
The manifold must be screwed counter-clockwise
into the handle so that it audibly snaps into place.
You must not pull the manifold downward.
Check the instrument for tightness, mobility and
accuracy (see page 63 "Servicing").
Note:
Servicing and Cleaning
67