Easy calibration – BrandTech Scientific Transferpette-8/-12 electronic User Manual

Page 60

Advertising
background image

1x

1x

1x

>3s

Press and hold the MENU
key (> 3 sec) until CAL
appears. The display reads
‚off‘. ‚CAL‘ blinks.

Press one of the arrow
buttons to activate the CAL
mode. The display changes
from ‚off‘ to ‚on‘. ‚CAL‘ con-
tinues to blink.

Press the ENTER key. The
display now shows the set
pipetting volume.
‚CAL‘ blinks.

Press the ENTER key.
The display shows the tested
and corrected volume. The
CAL symbol is continously
displayed to confirm that an
adjustment has been made.

Use the arrow keys (+/-) to
set the volume, which was
previously determined and
tested. ‚CAL‘ blinks.

The instrument should be set to either the nominal volume (for example
100 µl for a 100 µl pipette) or a specific test volume, in the standard
pipetting mode (PIP). See page 44 and 48 for procedures.
E.g., volume according to testing of volume 101,3 µl.

Easy Calibration

Adjustment of the Transferpette

®

-8/-12 electronic

1.

Bring up the CAL

mode

2.

Activate the CAL

mode

3.

Confirm CAL mode

4.

Set the volume

5.

Confirm volume

The calibration mode ‚CAL‘

Adjustment

What to do

How to do it

Keys to press

Display readout