IKA C 7000 Grundausstattung Set 2 User Manual

Page 25

Advertising
background image

IKA

-WERKE C 7000

Ver. 07 09.07

3DJH 

Almost all substances that are tested contain some sulphur and nitrogen. Under the

conditions employed for calorimetric measurements, sulphur and nitrogen burn to

SO

2

, SO

3

and NO

x

. In combination with water from combustion and moisture from

the sample, sulphuric acid and nitric acid are formed and heat of solution is devel-

oped. The influence of the heat of solution must be taken into account when calcu-

lating the standard calorific value.

To obtain a defined end condition, and to quantitatively record all acids, ca. 5 ml of

distilled water or other suitable absorption liquid are added to the decomposition

vessel. The gases resulting from combustion form acids with this absorption liquid

and any water resulting from combustion. After combustion, the decomposition ves-

sel is rinsed thoroughly with distilled water, to collect any condensate that has set-

tled on the walls of the vessel. The acid content of this solution can be determined

using suitable equipment.

You can obtain details of suitable equipment for the purpose from IKA

or from your

authorised dealer.

 &RPSOHWH &RPEXVWLRQ

To determine the gross calorific value correctly, it is of fundamental importance that

the sample has burnt completely. After a test, the crucible and all solid residues

must be examined for signs of incomplete combustion.

Normally, solids can be burnt directly in powder form. Substances that burn very fast

(e.g. benzoic acid) must not be burnt in a loose form. Such substances tend to ex-

plosive combustion, and complete combustion cannot then be guaranteed. In addi-

tion, the decomposition vessel could be damaged. These substances must be

pressed to form tablets using a special press (C 21 pelleting press, see Accesso-

ries) before testing.

Substances with low flammability (substances with a high mineral content or a low

calorific value) can often only be burnt completely with the aid of combustion cap-

sules or combustion bags (C 10/C 12, see Accessories). The use of liquids to aid

combustion, for example paraffin, is also possible.

Most liquids can be weighed directly in the crucible, but highly volatile substances

should be filled into combustion capsules (gelatine capsules or acetobutyrate cap-

sules, see Accessories) and burnt together with the capsules.

All combustion and ignition aids (e.g. cotton threads) must be burnt completely. If

anything remains unburnt, the test must be repeated.

Halogen-containing substances can cause corrosion of the decomposition vessel.

For such cases, decomposition vessel C 7012 should be used.

6ROLGV

6XEVWDQFHV

ZLWK ORZ

LQIODPPLELOLW\

+DORJHQV

$FLG FRUUHFWLRQ

/LTXLGV KLJKO\

YRODWLOH

VXEVWDQFHV

Advertising
This manual is related to the following products: