IKA C 7000 Grundausstattung Set 2 User Manual

Page 40

Advertising
background image

 3UHSDUDWLRQ DQG &DUU\LQJ 2XW 0HDVXUHPHQWV

IKA

-WERKE C 7000

Ver. 07 09.07

6HLWH 

g

0HDVXUHPHQW

In the C 7000 calorimeter, measurement is fully automatic.

h

&RROLQJ

The decomposition vessel is removed from the C 7000 calorimeter and placed in the

C 7002 cooler, where it is automatically cooled to the temperature of the calorimeter.

i

9HQW

The decomposition vessel is vented using the C 7010.8 vent handle or venting sta-

tion C 7030.

j

&OHDQ

Clean and check the decomposition vessel as described in Section 7.6.

2SWLPLVLQJ XVH RI WKH FDORULPHWHU

As can be seen in the diagram, a complete test procedure can be sub-divided into

several phases. To make maximum use of the calorimeter, you need four decompo-

sition vessels (DV No. 1 to DV No. 4 ) which are used as follows:

As one decomposition vessel is being prepared for a test (steps 1 to 4 in the cycle of

operations), a second vessel is undergoing a test (step 5 in the cycle of operations).

At the same time, a third decomposition vessel is cooling (step 6 in the cycle of op-

erations), while the fourth is being vented and cleaned (steps 7 and 8 ).

By using four vessels in parallel in this way, well over 100 tests can be carried out

with one calorimeter in a single day. By using two or three decomposition vessels in

a similar manner, a proportionately smaller time saving can be achieved.

To obtain the most accurate results, it is essential to observe the cycle of operations

shown above.

 5HFRPPHQGDWLRQV IRU FDOLEUDWLRQ

Before precise measurements can be made with the calorimeter, it must be cali-

brated. To calibrate the instrument, tablets of

FHUWLILHG EHQ]RLF DFLG

(see accesso-

ries) with a known calorific value are burnt. This test is used to determine the quan-

tity of heat required to raise the temperature of the calorimeter by 1 Kelvin. This is

known as the heat capacity (C value) of the system. This constant is needed later to

calculate gross calorific values.

The heat capacity is influenced both by the measuring cell and by the decomposition

vessel. ). It has a substantial influence on the gross calorific value being determined,

and must be re-established from time to time, particularly at commissioning, follow-

ing servicing and when parts are replaced.

Advertising
This manual is related to the following products: