The term "measurements" as used here means both measurements for calibrating

the instrument (calibration measurements) and measurements made to determine a

gross calorific value. The difference lies mainly in the evaluation of the results,

preparation and execution are almost identical for both types of measurement.

A complete test procedure comprises the following operations and procedures (for

detailed information about the individual steps, please see further sections in this

chapter):

c

3UHSDUH VDPSOH

For example, homogenise sample material, press to form pellets, dry

d

:HLJK

Weigh the sample in a crucible to an accuracy of 0.1 mg

e

3ODFH VDPSOH

Place the sample in the decomposition vessel and close the vessel .

f

)LOO ZLWK R[\JHQ

Fill the decomposition vessel manually with oxygen at 30 bar, quality 3.5, using the

IKA

C 48 oxygen filling station.

5.M easurem ent

6.C ooling

7.Vent

8.C lean

1.Prepare

sam ple

3.Place

sam ple

4.Fill with

oxygen

2. W eigh

'9 1R 

'9 1R 

'9 1R 

'9 1R 