IKA C 7000 Grundausstattung Set 2 User Manual

Page 51

Advertising
background image

IKA

-WERKE C 7000

Ver. 07 09.07

3DJH 

 &DUU\LQJ RXW D PHDVXUHPHQW

When a calibration or a calorific value determination has been prepared as de-

scribed above, the measuring process can be started. The

VDPSOH FRGH

shown on

the LCD-display informs you of the test parameters which will be used for the meas-

urement. Pressing the

button starts the calibration or calorific value determination.

From now on, the C 7000 calorimeter takes control. The measurement process is

made up of a pre-test and the main test. The current phase of the test is shown on

the LCD-display. During both the pre-test and the main test, you can follow the tem-

perature changes in the decomposition vessel on the LCD-display. From the start of

the test, they are updated every 1.5 seconds. In the second line of the LCD-display

you can also see the duration of pre-test or main test, which is indicated by an ex-

tending bar.

(QG RI WKH WHVW

The end of the test can be seen on the LCD-display or on the print out. As long as

the value



has not been entered in the sub-menu

UHDOW

(Real time printout), during

the test the printer will print the current temperatures at the set time intervals and the

result at the end of the test. In the first line, the LCD-display shows the temperature

increase in Kelvin and, for a calibration, the provisional C value in the second line.

The following steps must be carried out when the test has been completed:

c

The green LED blinking near the

9

button indicates the end of the test. Acknowl-

edge this by pressing the

9

button. If the button is not pressed, an acoustic signal

sounds after about five seconds, as an additional means of making you aware that

the test has ended. It sounds for about 30 seconds if it is not acknowledged. After

that, the LCD-display returns to

PHQX 

in the main menu.

d

Open the cover of the measuring cell by lifting the upper section of the instrument

and pushing it backwards. The result of the test just carried out then appears on the

LCD-display. Use the carrying and venting handle, C 7010.8, to lift the decomposi-

tion vessel out of the measuring cell.

e

So that the next test can take place under the same temperature conditions, the

energy absorbed by the decomposition vessel must be removed again. The C 7002

cooler fulfils this function. Place the decomposition vessel vertically in the cooler.

The cooling jaws close automatically. The display shows the temperature difference

between the decomposition vessel and the calorimeter in percent. During the cool-

ing process, this value progressively approaches zero. When the temperature of the

decomposition vessel is the same as the calorimeter temperature, this is indicated

by repeated, short acoustic signals. The cooling jaws open at the same time. Re-

move the decomposition vessel using the carrying and venting handle, C 7010.8.

The cooler can be manually controlled using the

RSHQFORVH

button. Pressing the

RSHQFORVH

button during the cooling process opens the cooling jaws. If the button is

operated again, the jaws close again and cooling is continued. Cooling only contin-

ues as long as the decomposition vessel is at a higher temperature than the calo-

rimeter.