IKA C 7000 Grundausstattung Set 2 User Manual

Page 42

Advertising
background image

 3UHSDUDWLRQ DQG &DUU\LQJ 2XW 0HDVXUHPHQWV

IKA

-WERKE C 7000

Ver. 07 09.07

6HLWH 

&RGLQJ

When working with the calorimeter, up to a maximum of 8 decomposition vessels

can be used. This is made possible by giving the decomposition vessels code num-

bers from 0 to 7. The calorimeter then recognises automatically which vessel is be-

ing used for a test, and selects the correct calibration parameter.

Coding is carried out by separating the narrow printed-circuit conductors on the

board mounted on the base of each decomposition vessel. The conductors are

marked I, II, and III. When separating the narrow conductors, it is vital to avoid dam-

aging the broad conductors. The conductors are separated at spacings of ca. 2 mm

using a sharp knife, and the intermediate segment removed.

1 3 2

The table below indicates which conductors must be separated for a particular code

number. For example: for code number 0, no conductors have to be separated, for

code number 3, conductors II and III must be separated.

Code number

Conductors to be separated

I

II

III

0

1

2

3

4

5

6

7

&RGLQJ

GHFRPSRVLWLRQ

YHVVHOV

1

Conductor I

2

Conductor II

3

Conductor III

Advertising
This manual is related to the following products: