Here the user also has the choice of manual entry or automatic input from a con-

nected balance. Manual entries are made directly in this window using the alpha-

numeric keys, or automatic input is called up by pressing function key

I

. The mass

weighed by the balance is automatically accepted. After entry, confirm by pressing

the

button. If you are preparing for a calibration, the LCD-display asks for the ref-

erence calorific value,

+VWDQG

If you are preparing for a calorific-value determina-

tion, the second line of the LCD-display briefly shows the enquiry

PHP VDYH

.

When it has disappeared, pressing the

PHP

button stores the data entered and

closes the entries for a calorific-value determination.

If you want to check the values before you store them, you can select each pa-

rameter using the arrow and function buttons and make corrections if necessary.

Once the parameters have been stored, it is no longer possible to change them.

If you are carrying out a calorific-value determination, omit the next step.

h

Only when preparing for a calibration: enquiry for reference calorific value,

+VWDQG

.

A charge-related pre-setting in the order of 26450 J/g will be displayed. This is the

calorific value of the benzoic acid tablets supplied by IKA

. If you are going to use a

different substance for calibration, use the key panel to enter the correct value; the

units must be J/g. Confirm by pressing the

button. The second line of the LCD-

display briefly shows the enquiry

PHP VDYH

. When it has disappeared, pressing

the

PHP

button stores the data entered and closes the entries for a calibration.

The message "Bomb securely closed?" will appear. Ensure that the decomposition

vessel is properly closed and confirm with OK.

i

You can now place the decomposition vessel in the calorimeter. Use the carrying

and venting handle, C 7010.8. Place it on the decomposition vessel and turn it anti-

clockwise until the handle engages. Take care not to accidentally press the vent

knob, which will cause oxygen to escape. Guide the decomposition vessel into the

measuring cell, until it is standing vertically and centrally in the cell. Remove the

carrying and venting handle by turning it clockwise. Close the cover by drawing the

complete upper section of the instrument over the measuring cell. The LCD-display

then shows the text

VDPSOH FRGH

and the parameters just entered.

Activate

6WDUW

Each time 1000 ignitions have been performed using a given decomposition vessel,

the following message will appear:

1000 ignitions performed with Bomb x

Inspect decomposition vessel

or contact IKA service

This indicates that the decomposition vessel has reached a maintenance point and

that a safety check must be carried out. Confirm this message by pressing

  7$%

and

2.

in sequence.

This message does not release the user from the responsibility of also continuously

checking the decomposition vessel for wear and carrying out safety inspections as

required.