3 basic operation, Basic operation – INFICON VGC403 Three-Channel Measurement and Control Unit User Manual

Page 20

Advertising
background image

20

tinb07e1-e 2011-07 Vacuum Gauge Controller

4.5.3

Basic operation

Starting at the measurement menu, you can select and
modify a specific parameter as follows:

1

Press the CH button to select the required channel.
(This is only necessary if you want to modify a sen-
sor parameter.)

2

Press the PARA button
• The parameter menu is selected
• The PARA indicator is illuminated

3

Use the arrow buttons to select the required
parameter group
• The name of the parameter group is displayed

4

Press the PARA button to select the required
parameter
• The name and the value of the parameter are

displayed

5

Use the arrow keys (and the CH button, if neces-
sary) to modify the parameter value
• The value of the parameter is changed

6

Repeat the steps 4 and 5 to change further param-
eters of the same parameter group

The unit returns to the measurement mode after the last
parameter of a parameter group has been accessed.
Parameter modifications are effective immediately, and
they are saved in the EEPROM automatically.

Advertising