3 channels, 1 sensor connections, 2 sensor supply – INFICON VGC403 Three-Channel Measurement and Control Unit User Manual

Page 8: 3 measuring technique, 4 switching functions, Channels, Sensor connections, Sensor supply, Measuring technique, Switching functions

Advertising
background image

8

tinb07e1-e 2011-07 Vacuum Gauge Controller

2.3 Channels

2.3.1

Sensor connections

2.3.2

Sensor supply

2.3.3

Measuring technique

NOTE:
The measurements of the BPG/BCG/HPG/CDGxxxD
are transferred digitally.

2.4 Switching functions

Number of channels

VGC402: 2
VGC403: 3

Sensor connections per
channel

RJ45 (FCC 68)
D-Sub, 15 pins, female
(connected in parallel)

Compatible sensors

Pirani:
PSG400, PSG400-S,
PSG100-S, PSG101-S,
PSG500, PSG500-S,
PSG502-S, PSG510-S,
PSG512-S, PSG550,
PSG552, PSG554

Pirani / Capacitance:
PCG400, PCG400-S,
PCG550, PCG552, PCG554

Cold cathode:
PEG100

Cold cathode / Pirani:
MPG400, MPG401

Capacitance:
CDG025, CDG025D,
CDG045, CDG045D,
CDG100, CDG100D,
CDG160D

Hot ionization / Pirani:
BPG400, BPG402
HPG400

Hot ionization / Pirani /
Capacitance: BCG450

Voltage

+24 VDC ±5 %

Current

500 mA
(750 mA short-time)

Fuse

900 mA via PTC element

Self-resetting after switching
the unit off or unplugging the
sensor.

The supply meets the
requirements of a ground
protective extra low voltage
(SELV).

Measuring ranges

Sensor dependent

Error of measurement

Gain error:

≤ 0.005 % FS
Offset error:

≤ 0.01 % FS

Measuring rate

50 s

-1

Display rate

10 s

-1

Filter time constant

Slow:
Approx. 1.0 s (f

g

= 0.16 Hz)

Normal (nor):
Approx. 0.3 s (f

g

= 0.53 Hz)

Fast:
Approx. 0.06 s (f

g

= 2.65 Hz)

Unit of measurement

mbar, Pa, Torr, Micron

Possible adjustments

Linear sensors (CDG):
Zero-adjust

Logarithmic sensors (PSG,
PCG, PEG, MPG, BPG,
BCG, HPG):
Fixed correction factors for
N

2

, Ar, H

2

, or a variable cor-

rection factor in the range
0.10…9.99

A/D converter

Resolution > 16 bit

Number of switching
functions

VGC402: 4
VGC403: 6

Assignment

Can be configured any way

Delay time

Filter time constant depen-
dent

Adjustment range

Sensor dependent

Hysteresis

Linear sensors (CDG):

≥ 1

% FS

Logarithmic sensors (PSG,
PCG, PEG, MPG, BPG,
BCG, HPG):

≥ 10 % of measurement

Advertising
This manual is related to the following products: