2 configuring switching functions, 3 adjustment range, Configuring switching functions – INFICON VGC403 Three-Channel Measurement and Control Unit User Manual

Page 22: Adjustment range

Advertising
background image

22

tinb07e1-e 2011-07 Vacuum Gauge Controller

5.1.2

Configuring switching functions

Prerequisite: The parameter group SP-P is selected

1

Press the PARA button to select the required
parameter
• The name and the value of the parameter are

displayed

2

Use the CH button to assign the switching function
to a channel
• The switching functions can be assigned to the

channels any way

• The two threshold values of the switching func-

tion are always assigned to the same channel

3

Use the arrow buttons to modify the threshold
value
• The value of the parameter is changed

4

Repeat the steps 1 to 3 to change further parame-
ters of the same parameter group

5.1.3

Adjustment range

Adjustment range of the lower threshold value
The lower threshold value of a switching function can be
set in the following pressure range:

Tab. 5-1

Adjustment range of the lower threshold values

*) 2

×

10

-4

mbar if

PrE

is activated (see Chapter 5.2.12,



26)

Adjustment range of the upper threshold value
The upper threshold value of a switching function can be
set in the following pressure ranges:

Tab. 5-2

Adjustment range of the upper threshold values

This means that the hysteresis amounts to 10 % of the
lower threshold value (logarithmic sensors) or to 1 % of
the full-scale range (linear sensors) at least. If another
sensor type is connected to a channel, the respective
threshold values will be adjusted automatically if neces-
sary.

Sensor

SP-L

min.

[mbar]

SP-L

max.

[mbar]

PSG

2

Ч 10

-3

*)

5

Ч 10

2

PCG

2

Ч 10

-3

*)

1.5

Ч 10

3

PEG

1

Ч 10

-9

1

Ч 10

-2

MPG

5

Ч 10

-9

5

Ч 10

2

CDG/CDG d

FS/1000

FS

bPG/bPG2

1

Ч 10

-8

5

Ч 10

2

HPG

1

Ч 10

-6

1

Ч 10

3

bCG

1

Ч 10

-8

1.5

Ч 10

3

Sensor

SP-H

min.

[mbar]

SP-H

max.

[mbar]

PSG

1.1 SP-L

5

× 10

2

PCG

1.1 SP-L

1.5

× 10

3

PEG

1.1 SP-L

1

× 10

-2

MPG

1.1 SP-L

5

× 10

2

CDG/CDG d

SP-L

+ 0.01 FS

FS

bPG/bPG2

1.1 SP-L

5

× 10

2

HPG

1.1 SP-L

1

× 10

3

bCG

1.1 SP-L

1.5

Ч 10

3

Advertising