2 mains connection, 3 ground, 4 sensor – INFICON VGC403 Three-Channel Measurement and Control Unit User Manual

Page 12: Mains connection, Ground, Sensor, Warning

Advertising
background image

12

tinb07e1-e 2011-07 Vacuum Gauge Controller

3.3.2

Mains connection

The mains connection (Fig. 3-4,  11, Pos. G) is
designed for a mains cable which contains a European
appliance connector on the device side.

A mains cable is supplied with the unit. If the plug is not
compatible with your wall socket, you have to get a suit-
able mains cable:

• Three-conductor cable with protective ground
• Conductor cross-section 3

× 1.5 mm

2

or larger

Fig. 3-5

Three-conductor cable with protective ground (example)

1

Connect the European appliance connector of the
mains cord with the mains connection of the unit

2

Connect the plug of the mains cable with the wall
socket

NOTE:
If the unit is installed in a switching cabinet, the
mains power can be supplied via a switchable central
power distributor.

3.3.3

Ground

The ground screw (Fig. 3-4,  11, Pos. I) can be used to
connect the Vacuum Gauge Controller with the protective
ground of the pumping station.

1

If required: Connect the protective ground of the
pumping station with the ground screw. Use a pro-
tective conductor.

3.3.4

SENSOR

The SENSOR connection is used to connect the sensors.

For each channel, there are two connections available
which are connected in parallel: An 8-pin RJ45 appliance
socket and a 15-pin D-Sub appliance socket. See
Fig. 3-4,  11, Pos. A…C.

Pin assignment

Fig. 3-6

SENSOR appliance socket (RJ45)

Fig. 3-7

SENSOR appliance socket (D-Sub, 15-pin)

WARNING

Mains power.

Improperly grounded devices can be ex-
tremely dangerous in the event of a fault.

Use three-wire mains or extension cables
with protective ground only. Plug the mains
cable into wall sockets with protective ground
only.

WARNING

No mains fuse.

The Vacuum Gauge Controller is not
equipped with a mains fuse.

The wall socket must be protected with a fuse
(max. 10 A).

1 +24 VDC
2 PGND
3 Signal
4 Ident

5 Signal-GND
6 Status
7 HV_L
8 HV_EMI

1 EMI-Status
2 Signal
3 Status
4 HV_EMI
5 PGND
6 n.c.
7 Degas
8 Supply

9 n.c.

10 Ident
11 Supply_CDG
12 Signal-GND
13 RXD
14 TXD
15 Chassis

1

8

1

8

9

15

Advertising
This manual is related to the following products: