6 control, 7 rs232c, Control – INFICON VGC403 Three-Channel Measurement and Control Unit User Manual

Page 14: Rs232c

Advertising
background image

14

tinb07e1-e 2011-07 Vacuum Gauge Controller

3.3.6

CONTROL

The CONTROL connection (Fig. 3-4,  11, Pos. D) con-
tains the following signal pins:

• Analog outputs for the signals of the individual chan-

nels

• Recorder output. This is a programmable analog out-

put which can be assigned to one of the three chan-
nels.

• HV-EMI. Used to switch the high-vacuum circuit of the

PEG sensor on and off. The signal levels are: On =
+24 V. Off = 0 V. See Reference [7].

Pin assignment

Fig. 3-9

CONTROL appliance plug (D-Sub, 9-pin)

1

Connect the peripheral components with the CON-
TROL connection. Use a shielded connection
cable.

NOTE:
The analog outputs (pins 1, 2, 6) differ from the dis-
played values by no more than ±50 mV.

3.3.7

RS232C

The RS232C serial interface (Fig. 3-4,  11, Pos. F)
allows remote control of the unit via a computer or a ter-
minal. See Chapter 6 Computer interface,  32.

In addition, the interface may be used for firmware
updates. See Chapter 7.2 Program transfer mode,  45.

Pin assignment

Fig. 3-10

RS232C appliance socket (D-Sub, 9-pin)

1

Connect the serial interface of the computer with
the RS232C connection. Use a shielded cable.

NOTE:
Use a serial extension cable with a 9-pin plug and a
9-pin socket. The cable must not contain any crossed
wires.

1 Analog output 1
2 Analog output 3
3 GND
4 HV-EMI 3
5 HV-EMI 1

6 Analog output 2
7 Recorder output
8 GND
9 HV-EMI 2

1

5

6

9

1 n.c. / SUP
2 TXD
3 RXD
4 n.c.
5 GND

6 DSR
7 n.c.
8 CTS
9 GND

1

5

6

9

Advertising